316. INTRAOPERATIVE CONTROL OF OSTEOSYNTHESIS WITH A 3D AMPLIFIER: ONE YEAR’S EXPERIENCE



Abstract

Purpose of the study: Computed tomography is recommended for the preoperative work-up of joint fractures as it allows an optimisation of the access as a function of the injury. During the operation, 2D radiographic or fluoroscopic controls are still widely used. After one year’s experience, we evaluated the potential pertinence of using 3D reconstructions intraoperatively with a mobile isocentric fluoroscope (iso-C-3D).

Material and methods: All operations for which the amplifier was used were collected prospectively. The type of fixation as well as the details of the installation and measures taken intraoperatively were noted.

Results: At one year, intraoperative 3D reconstructions were made during 48 operations in 47 patients: fracture of the calcaneum (n=13), thoracolumbar spin (n=12), acetabulum (n=11), tibial condyles (n=9), odontoid (n=2), pelvis (n=1). The installation was habitual for the calcaneum and odontoid fractures. For the other localizations, use of a carbon plateau table facilitated good quality imaging for spinal and tibial condyle fractures; a carbon orthopaedic table was useful for acetabulum and pelvis fractures. With the intraoperative 3D reconstruction the surgeon was able to check the freedom of the canal after reduction and fixation. For the calcaneum fractures, reduction of the thalamic fragment was revised in one patient; in another, an intra-articular screw was replaced. One intra-articular screw stabilizing the posterior wall was also changed during an acetabulum fixation.

Discussion: During our first year of use, 3D reconstruction intraoperatively has allowed us to avoid three early reoperations (for two calcaneums and one acetabulum). Classical 2D imaging of these two localizations is difficult to interpret because of the spherical form of the hip joint and, for the calcaneum, the difficulty in obtaining quality retrotibial images. Quality images requires specific installation, limiting interference with metallic supports.

Conclusion: The results we have obtained in our first year of use of the ISO-C-3D amplifier has led us to generalise its use for percutaneous fixation procedures involving the acetabulum and the calcaneum.