280. MEASURING INSERT WEAR OF A TOTAL KNEE ARTHROPLASTY: ALGORITHMIC APPROACH USING RADIOSTEREOMETRIC ANALYSIS (RSA)



Abstract

Purpose of the study: Several authors have focused attention on the kinematics of total knee arthroplasty. In vitro studies have shown the influence of prosthetic kinematics on wear of the tibial insert. In vivo, investigations are limited for this important parameter. We propose computer algorithm which gives a linear assessment of femur penetration into the tibial insert. The purpose of this work was to test this measurement by radiostereometric analysis (RSA) on pre-worn prosthesis models.

Material and methods: A resin plateau was manufactured with four Triatlhon posterior stabilized inserts. Three of these inserts were pre-worn on the medial glenoid; wear varied from 0.8 to 1.2 mm. The wear pattern was controlled by a computer program in order to localize it on a posteromedial gliding band posteriorly to the posterior stabilization implant. A femoral component was implanted on dry bone and loaded on a simulator. RSA images were obtained for each of the plateau. Five series of images were obtained from 0 to 40° flexion. The measurement method consisted in defined a perfect middle plane based on the analysis of tantalum beads embedded in the resin plateau, the to define the point the closest to this plane belonging to the medial condyle. The norm of the normal vector for this plane passing through the lowest point of the femur was the reference. The difference of the norms of the vectors constructed from the healthy plateau and the worn plateaus was defined as penetration of the femur into the tibia insert. The rood mean square (RMS) method was used to measure intraobserver variability and the interclass coefficient of correlation (ICC) was determined.

Results and Discussion: Our results were encouraging. 0.8/to 1.12 mm wear was detected with very good accuracy. The data spread was however wide, irrespective of the wear value. The images in flexion of this model provided the most reliable readings. This study is the first offering an approach to wear measurement in vivo. The data spread should be improved by fine tuning the experimental model and the pre-wear pattern.

Conclusion: In vivo measurement of TKA wear remains a challenge in clinical medicine. Our work demonstrate the feasibility of using RSA.