header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

259. RELIABILITY OF WEAR MEASUREMENTS FOR THE DUAL MOBILITY CUP: EXPERIMENTAL RADIOSTEREOMETRIC ANALYSIS (RSA)



Abstract

Purpose of the study: The concept introduced by Gilles Bousquet is an effective arm against dislocation of total hip arthroplasty (THA), as has been demonstrated in clinical series with a long follow-up. There remain certain questions concerning wear of dual mobility cups. We propose a radiostereometrical analysis (RSA) of femoral head migration in this type of implant. Our objective was to establish an accurate measurement and determine the intra- and interobserver variabilities.

Material and methods: A THA model was implanted and loaded with a simulator. Penetration of the implants was measured using a specially designed polyethylene insert with increasingly concentric wear (from 0, 0.25, 0.5 to 0.75 mm). Three investigators analysed (7 times in a double-blind protocol) the RSA images of these four inserts. The investigators were an expert (I), well-trained (II), naive (III). The accuracy of the measurement as well as the intra- and interobserver variabilities were determined using the root mean square (RMS) method, the interclass coefficient of correlation (ICC), the Bland and Altman analyses, and weighted Kappa analysis.

Results: Regarding accuracy, the RMS was 0.0388 [CI95: 0.02266–0.05564]. The mean error for preworn inserts was respectively 0.022mm (for 0.25mm prewear), 0.59mm (for 0.5mm), and 0.022mm (for 0.75mm). The intra-observer ICC was 0.9714 [0.9028–0.9918] for investigator I. The interobserver ICCs between investigators I and II and between I and III were respectively 0.943 and 0.968. The weighted kappa coefficients between I and II and between I and III were 0.827 and 0.849. The Bland and Altman analysis confirmed these results.

Discussion: Several RSA protocols could be designed to measure wear of the dual mobility cup. We chose detection of the wear pattern instead of the tantalum beads method. Our protocol, using a simple geometric model and not the manufacturers CAD files, showed an accuracy comparable with manufacturing tolerances with low variability.

Conclusion: This study validated our measurement method, a prerequisite for a randomized multicentric study which has been initiated to compare, by RSA, penetration of the head into the double mobility insert versus a fixed insert.

Correspondence should be addressed to Ghislaine Patte at sofcot@sofcot.fr