header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

A830. IN VITRO LYMPHOCYTE ASSAY FOR HYPERSENSIVITY TO METALS DURING THE FIRST TWO POSTOPERATIVE YEARS USING TWO DIFFERENT RESURFACING HIP ARTHROPLASTIES



Abstract

A major concern in metal on metal bearings has been the elevated serum concentrations of cobalt and chromium. Recent papers have suggested that metal hypersensivity in a few cases could cause periprostetic lymphocyte accumulation leading prosthetic loosening.

To measure the lymphocyte activation and proliferation in vitro by re-exposure of the cells to cobalt, chromium, nickel and titanium. To correlate the lymphocyte assay data to the serum concentration of metals and plasma cytokines.

A prospective clinical study with the ASR (DePuy) and ReCap (Biomet) resurfacing hip implants. Blood samples were collected one and two years postoperatively, lymphocytes were isolated by density gradient centrifugation, cultured in a medium containing the patient’s serum and exposed to metal salts.

Cells were analyzed by flow cytometry, evaluating number, viability, size and CD69 activation.

A negative control and a positive control (phytohaemagglutinine) were included in the assay, and the responses to the metals were calculated in proportion to controls. 11 patients were assessed at one and two years follow up, 16 patients were assessed only at two years.

Serum chromium and cobalt were measured preoperatively, six months, one year and two years postoperatively by graphite furnace absorptiometry. Plasma cytokines were measured by multiplexed immunoassay.

In the assay the negative and positive controls gave the expected responses.

When exposed to metals no response was found in the lymphocytes in any patients.

There were no difference in response between one and two years.

The results seems to indicate that the metal hypersensitivity is a rare condition in metal on metal arthroplasty. The results indicate that the method can be used to monitor hypersensitivity to implant metals.

Correspondence should be addressed to Diane Przepiorski at ISTA, PO Box 6564, Auburn, CA 95604, USA. Phone: +1 916-454-9884; Fax: +1 916-454-9882; E-mail: ista@pacbell.net