header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

A1038. ACCURACY OF ROBOTICALLY ASSISTED UKA



Abstract

Successful clinical outcomes following unicompartmental knee arthroplasty (UKA) depend on accurate component alignment, which can be difficult to achieve using manual instrumentation. A new technology has been developed using haptic robotics that replaces traditional UKA instrumentation. This study compares the accuracy of UKA component placement with traditional jig-based instrumentation versus robotic guidance.

85 UKAs performed using standard manual instrumentation were compared to 67 performed with a robotically guided implantation system without instrumentation. Each was performed using a minimally invasive surgical approach. The two groups were identical in terms of age, gender and BMI. The coronal and sagittal alignment of the tibial components were measured on pre-and post-operative AP and lateral radiographs. Postoperative tibial component alignment was compared to the pre-operative plan.

The RMS error of the tibial slope was 3.7° manually compared to 1.2° robotically. In addition, the variance using manual instruments was 9.8 times greater than the robotically guided implantations (p< 0.0001). In the coronal plane, the average error was 3.0 ± 2.2° more varus using manual instruments compared to 0.3 ± 1.9° when implanted robotically (p< 0.0001), while the varus/valgus RMS error was 3.7° manually compared to 1.8° robotically. The average depth of medial tibial plateau resection was significantly less with inlay tibial components (3.9 ± 0.9mm) relative to onlay tibial components (6.8 ± 0.9mm, p< 0.0001). In addition, a significantly higher percentage of robotic inlay patients went home the day of surgery (12% vs. 1%, p< 0.0001).

Tibial component alignment in UKA is significantly more accurate and less variable using robotic guidance compared to manual, jig-based instrumentation. By enhancing component alignment, this novel technique provides a potential method for improving outcomes in UKA patients.

Correspondence should be addressed to Diane Przepiorski at ISTA, PO Box 6564, Auburn, CA 95604, USA. Phone: +1 916-454-9884; Fax: +1 916-454-9882; E-mail: ista@pacbell.net