A910. METAL-ON-METAL HIP RESURFACING IN YOUNG PATIENTS: THE ROLE OF FEMOROACETABULAR IMPINEGMENT



Abstract

Femoroacetabular impingement (FAI) has been identified as the cause of idiopathic osteoarthritis in young patients. FAI is the result of decreased femoral head/neck offset ratio due to bony deformities and causes hip pain and labral tears. Because the unique design and bone preserving nature of metal-onmetal hip resurfacing implants, it is extremely difficult to correct extensive bony deformities associated with FAI. Poor patient selection and lack of orrection/undercorrection of the underlying FAI deformity may lead to prosthetic impingement, extensive wear and metal ion release, component loosening, and subsequent implant failure. Hence, it is critical to define the patient population undergoing hip resurfacing. Because metal-on-metal hip resurfacing is performed more frequently in a younger population, we hypothesize that this patient population will have a larger proportion of femoroacetabular impingement than the general population and identification of this patient population is critical to the longevity of the implant.

A retrospective review of 153 hips undergoing metal-on-metal hip resurfacing was performed. 52 hips were excluded based on the exclusion criteria of inadequate preoperative films (6 subjects), existing hardware/history of trauma (11 subjects), or if the resurfacing was performed due to avascular necrosis secondary to trauma, steroids, etc (35 subjects). The remaining 101 hips (76 male, 25 female) had an average age of 51.8 years. Preoperative x-rays were utilized to assess impingement according to previously published methods. An acetabular index (AI) of x ≤ 0°, center edge angle (CE angle) of x > 39°, a Sharp angle of x < 33°, and a present cross-over sign were considered pathologic findings for pincer impingement. Pathologic findings for cam impingement included the triangular index (TI; pathologic with R=r+2mm) and an α angle greater than 83° in men or 57° in women. Subjects were categorized as having impingement if they had one or more pathologic finding for either cam or pincer impingement and as having mixed impingement if they had one pathologic finding for both cam and pincer measurements. Prevalence rates were compared to published data for the general population.

Fifty-five subjects had at least one pathologic finding for cam impingement (18, 7, and 30 subjects had pathologic measurements for α angle, TI, and both measurements, respectively); 24 subjects had at least one pathologic pincer measurement (4, 6, 14, and 4 pathologic measurements for AI, CE angle, cross-over sign, and Sharp angle, respectively; 3 subjects had multiple pincer findings) 13 subjects were classified as having mixed impingement (with α angle and cross-over sign as the most prevalent cam and pincer measurements). When compared to published data for the general population (M: 17%, F: 4%), we found a significantly larger proportion of cam impingement in both males (60.5%) and females (36%) in patients undergoing resurfacing at our institution (p< 0.001). There was also a significantly larger proportion of pincer impingement in our population (23.8%) than in the general population (10.7%) (p=0.01). There was no significant difference between our proportion of patients with mixed impingement (12.9%) and the general population (20.8%) (p=0.150).

The patient population for metal-on-metal hip resurfacing shows a greater prevalance of FAI than the general population. Because the femoral head/neck junction is preserved with hip resurfacing, patients undergoing this type of procedure might be at increased risk of impingement. Hence, it is important to assess the degree of FAI preoperatively. This will allow proper patient selection and careful planning of surgical correction of the underlying FAI deformity to increase implant longevity.

Correspondence should be addressed to Diane Przepiorski at ISTA, PO Box 6564, Auburn, CA 95604, USA. Phone: +1 916-454-9884; Fax: +1 916-454-9882; E-mail: ista@pacbell.net