AB 1156. HIP RESURFACING ARTHROPLASTY: THE EFFECT OF ANTERIOR AND POSTERIOR NOTCHING ON FRACTURE RESISTANCE



Abstract

Surgeons performing hip resurfacing ante-vert and translate the femoral component anterior to maximize head/neck offset and educe impingement. The anterior femoral neck is under tensile forces during gait similarly to the superior neck [6]. This study was esigned to determine the risk of femoral neck fracture after anterior or posterior notching of the femoral neck.

Method: Fortyseven 4th generation synthetic femora were implanted with Birmingham Hip Resurfacing pros-theses (Smith & Nephew Inc. emphis, USA). Implant preparation was performed using imageless computer navigation (VectorVision SR 1.0, BrainLAB, Grmany). The virtual prosthesis was initially planned for neutral version and translated anterior, or posterior, to create the notch. The femora were fixed in a single-leg stance and tested with axial compression using a mechanical testing machine. This method enabled comparison with previously published data. The synthetic femora were prepared in 8 experimental groups:2mm and 5mm anterior notches, 2mm and 5mm posterior notches, neutral alignment with no notching (control), 5mm superior notch, 5mm anterior notch tested with the femur in 25° flexion and 5mm posterior notch tested with the femur in 25° extension We tested the femora flexed at 25° flex-ion to simulate loading as seen during stair ascent. [3] The posterior 5mm notched femoral necks were tested in extension to simulate sporting activities like running. The results were compared to the control group in neutral alignment using a one-way ANOVA:

Results: Testing Group Mean load to failure Significance Neutral (Control) 4303.09 ± 911.04N Anterior 2mm 3926.62 ± 894.17N p=0.985 Anterior 5mm 3374.64 ± 345.65N p=0.379 Posterior 2mm 4208.09 ± 1079.81N p=1.0 Posterior 5mm 3988.07 ± 728.59N p=0.995 Superior 5mm 2423.07 ± 424.16N p=0.003 Anterior 5mm in 25° flexion 3048.11 ±509.24N p=0.087 Posterior 5mm in 25° extension 3104.61±592.67N p=0.117 Both the anterior 5mm notch tested in single-leg stance and anterior notch in flexion displayed lower compressive loads to failure (3374.64N and 3048.11N). The mean load to failure value for the posterior 5mm notches in extension was 3104.62N compared to 4303.09N for the control group. Our data suggests that anterior and posterior 2mm notches are not statistically significantly weaker in axial compression. The anterior 5mm notches tend towards significance in axial compression (p=0.38) and bordered significance in flexion (p=0.087). The 5mm posterior notches were not significantly weakened in axial compression (p=0.995), but tended towards significance in extension (p=0.117). The 5mm superior notch group was significantly weaker with axial compression supporting previous data published (p=0.003). We are currently assessing offset and other variables that may reduce data spread.

Conclusion: We conclude that anterior and posterior 2mm notching of the femoral neck has no clinical implications, however 5mm anterior notches may lead to fracture. The fracture is more likely to occur with stair ascent rather than normal walking. Posterior 5mm notches are not likely to fracture with normal gait, but may fracture with higher impact activities that promote weight bearing in extension. Hip resurfacing is commonly performed on active patients and ultimately 5mm notching in the anterior or posterior cortices has clinically important implications.