079 BIOLOGICAL FIXATION IN DISTAL FEMORAL ENDOPROSTHESES: COMPARISON BETWEEN UNCEMENTED STEM AND THE COMPRESS® IMPLANT



Abstract

The objective of this study was to compare the results of two consecutive series of patients with either intra-medullary uncemented stems (UCS) distal femoral endoprosthetic replacement or the Compress® (CMP) distal femoral implant.

Patients were divided into two groups: those who received UCS prosthesis (Group-1: 54 patients) and those who received CMP prosthesis (Group-2: 42 patients).The most frequent diagnosis was osteosarcoma. Age and gender were similar both groups.

In Group-1, at a mean follow-up of 144 months, 37 prostheses were still in place. The overall Kaplan-Meier prosthetic survival rates were 79% at five and 62 % at ten years. Most of failures were long term complications. Aseptic loosening was the primary cause of late prosthetic failure. On Cox regression analysis, prosthetic stem diameter under 13mm was a significant negative prognostic factor for prosthetic survival (p=0.016).

In Group-2, at a mean follow-up of 84 months, 36 prostheses were still in place. The overall rate of CMP prosthesis survival was 86% at 5 years. All complications were during the first postoperative year, being femoral fracture the main revision cause.

The patients who retained the prosthesis had mainly good or excellent MSTS functional results in both groups.

Correspondence should be addressed to Anastasia C. Tilentzoglou MD, General Secretary of the Board of Directors of HAOST, 20 A. Fleming Str. (N.Filothei), Gr. 15123 Maroussi, Athens Greece. E-mail: info@eexot.gr