header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

147. A PROSPECTIVE RANDOMISED CLINICAL TRIAL: DOES HEAD SIZE EFFECT METAL ION LEVELS IN METAL-ON-METAL THAS?



Abstract

Purpose: Metal-on-metal articulations are an increasingly popular choice as an alternate bearing surface in total hip arthroplasty (THA) and Resurfacing implants. One advantage of a metal-on-metal bearing is the use of larger diameter femoral heads with hip simulator data demonstrating reduced wear. We performed a prospective, multicentre, randomized, blinded clinical trial comparing 28mm to 36mm metal-on-metal bearings assessing multiple validated outcome measures and serum, erythrocyte and urine metal ions.

Method: Ninety-one patients were randomized to receive a metal-on-polyethylene (34), a 28mm metal-on-metal (25) or a 36mm metal-on-metal (32) insert. All patients received the same acetabular and femoral component. Patients were evaluated pre-operatively, at 6, 12 months and annually thereafter, including an evaluation of serum, erythrocyte and urine cobalt, chromium, and titanium, outcome measures (WOMAC, SF-12, Harris Hip Score) and radiographs.

Results: At a minimum two years follow-up there were no differences in WOMAC, SF-12, Harris Hip scores or radiographs. Patients receiving metal liners had significantly (p< 0.001) elevated metal ion measurements compared with the polyethylene control group, however there were no differences between the 28mm and 36mm metal-on-metal bearings (Median serum Co (mg/L): 0.14(poly), 0.77(28mm), 0.73(36mm). Median erythrocyte Co (mg/L): 0.11(poly), 0.42(28mm), 0.42(36mm). Median urine Co(mg/day): 0.44(poly), 4.55(28mm), 5.42(36mm)). (Median serum Cr(mg/L): 0.17(poly), 1.29(28mm), 0.91(36mm). Median erythrocyte Cr(mg/ L): 1.10(poly), 1.10(28mm), 1.20(36mm). Median urine Cr(mg/day): 0.27(poly), 1.92(28mm), 2.02(36mm)).

Conclusion: Both cobalt and chromium ion measurements were significantly elevated in the blood and urine of the patients randomized to receive the metal-on-metal bearings at all time intervals. There were no differences seen between the 28mm and 36mm metal-on-metal bearings, keeping all other variables identical. The larger diameter bearing therefore provides the potential clinical advantages of improved range of motion and stability, while providing a similar metal ion profile. While reduced wear is seen with larger diameter metal-on-metal bearings in-vitro, we could not demonstrate a reduction in blood or urine metal ion levels in-vivo.

Correspondence should be addressed to CEO Doug C. Thomson. Email: doug@canorth.org