52. TOTAL HIP REPLACEMENT SURGERY UTILIZING THE ECHELON PRIMARY FEMORAL STEM



Abstract

Purpose: While the durability of most uncemented femoral stems remains unknown, it is the aim of this study to demonstrate Echelon Primary femoral stem performance with regard patient outcome and overall implant survival.

Method: Between February 1998 and March 2007, 428 patients received the Echelon Primary stem. The mean age of each patient was 58.1 (SD 11.1, Range 20–87). Body mass index averaged 30.5 kg/m2 (SD 5.8, Range 17.7–58.2). The majority of patients received a Reflection uncemented acetabular component (91%) and an ultra high molecular weight polyethylene liner (76.5%), although the highly cross linked polyethylene is now used with increased frequency, used in 31% of hips since 2005. The majority of femoral heads were cobalt chrome (79.3%).

Results: Kaplan Meier survivorship for the Echelon Femoral stem with revision for aseptic loosening as end point at 100 months is 99.3% (95% CI 97.1–99.8). Taking revision for any reason as the end point the Kaplan Meier survivorship is 98.3% at 100 months (95% CI 95.9–99.3). A pre-operative WOMAC score was available for 345 of the 392 patients with mean score of 43.5 (95% CI 41.6–45.4). At the three-month post-operative review the mean WOMAC score was significantly increased to 74.54 (95% CI 72.7–76.3)(p< 0.001) and by 1 year 84.3 (95% CI 80.5–88.1). At subsequent years, the modified WOMAC score remained at a plateau of around 80. General health assessment using the SF-36 shows an improvement in the physical component score from 33.1 (95% CI 32.3–33.9) preoperatively to 42.6 (95% CI 41.7–43.6)(p< 0.001) at three months and 48.19 (95% CI 44.2–52.2) at latest follow up. The mental component scores increased from 48.7 (95% CI 47.6–49.9) to 51.4 (95% CI 50.3–52.4)(p< 0.001) and 53.5 (95% CI 50.3–56.6) respectively.

Conclusion: This large prospective review of the Echelon Primary femoral stems reveals an excellent survivorship of the stem with a 99.3% survival at 8 years with regard aseptic loosening and 98.3% survival including revision for any reason. Patient outcome scores are significantly improved and subsequently maintained. There have been no changes with regard to manufacture or design of the stem within the period of review.

Correspondence should be addressed to CEO Doug C. Thomson. Email: doug@canorth.org