Full Access

22. OXIDATIVE STRESS MARKERS IN THE BLOOD OF PATIENTS WITH ARTICULAR SURFACE REPLACEMENT



Abstract

Purpose: The presence of metal ions in the blood of patients with a metal-on-metal (MM) bearing points to the importance of understanding the long-term effects of these ions. Metal ions have the potential to induce the production of reactive oxygen species (ROS), making them prime suspects for inducing molecular damage in circulating cells. The aim of this study was to analyze the levels of oxidative stress markers in the plasma of patients with hip surface replacement.

Method: Blood was collected up to 3 years after implantation from 66 patients with articular surface replacement (ASRÔ, DePuy Orthopaedics) and 54 patients with 36 mm-head MM THA. Forty (40) pre-operative patients were also assessed as control group. Total anti-oxidant levels were measured by the Oxford Biomedical total antioxidant power assay (Oxford, MI) to obtain an overview of the defense capacity of patient’s oxidative stress. Peroxide concentrations were measured by the Biomedica OxyStat assay (Medicorp, Montreal, QC) to quantify damage to lipids in the systemic circulation. Nitrototyrosine levels were quantified using the BIOXYTECH® Nitrotyrosine-EIA assay (OxisResearch™, Portland OR) to measure damage to proteins. The concentrations of metal ions were analyzed by inductively coupled plasma-mass spectroscopy.

Results: Results showed that there were no statistical differences in the concentrations of total antioxidants, lipid peroxides, and protein nitrotyrosines between the control, the ASR, and the 36 mm-head groups. Furthermore, there was no correlation between the concentrations of these markers and the concentrations of both Co and Cr ions (r2 £ 0.006).

Conclusion: The single most significant obstacle preventing a broader application of metal-on-metal hip arthroplasties and resurfacings continues to be the concerns regarding elevated metal ion levels in the blood and urine of patients. The present results showed that there were no changes in the levels of oxidative stress markers in patients with MM bearings compared to the control group. Given the possible latency periods related to metal ion exposure, longer follow-ups are required to conclusively determine the effects of elevated circulating ions on oxidative stress in the blood of patients with MM bearings.

Correspondence should be addressed to CEO Doug C. Thomson. Email: doug@canorth.org