header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

NOVEL CERAMIC-ON-METAL HIP REPLACEMENTS



Abstract

Ceramic-on-metal (ceramic head and metal liner, COM) hip replacements have shown reduced wear in comparison to metal-on-metal (MOM) bearings. This has been attributed to reduced corrosive and adhesive wear, and differential hardness. The study assessed the performance of ceramic and metal bearings in different configurations under adverse conditions, ceramic heads on metal liners (COM) were compared to metal heads on ceramic inserts (MOC), with head on cup rim loading under micro-separation hip joint simulation.

Components used were made of zirconia-platelet toughened alumina (Biolox Delta) and CoCrMo alloy. Hip simulator testing applied a twin-peak loading cycle and walking motions with the prosthesis in the anatomical position. Testing was conducted in calf-serum for 2-million cycles. A standard simulator cycle was adapted, the head sub-luxed in the swing-phase forcing the head onto the cup rim at heel strike.

The overall mean wear rate for the MOC bearings (0.71±0.30mm3/Mc) was significantly higher than the wear rate for the COM bearings (0.09±0.025mm3/Mc). The contact of the head against the rim of the cup caused deep stripe wear on the metallic heads of the MOC bearings. This region on the head is exposed to high stress conditions and susceptible to damage in edge contact, the effect of this is increased when the cup is a harder material than the head. The wear of a metal-on-metal (MOM) couple under similar conditions was almost two-fold greater than the MOC couple (1.58mm3/Mc, Williams et al., 2006) providing further evidence of the reduced wear with COM in comparison to MOM.

The COM concept allows thin metal shells to be used with larger ceramic heads and protects against ceramic liner chipping. COM bearings are undergoing clinical trials, early data suggests reduced metal ion release in patients compared to metal-on-metal.

Correspondence should be addressed to Editorial Secretary Mr ML Costa or Assistant Editorial Secretary Mr B.J. Ollivere at BOA, 35–43 Lincoln’s Inn Fields, London WC2A 3PE, England; Email: mattcosta@hotmail.com or ben@ollivere.co.uk