HIP-SPINE RELATIONSHIPS: BASIC PRINCIPLES, RADIOLOGICAL EXPLORATION AND DIRECT APPLICATIONS FOR TOTAL HIP REPLACEMENTS (RANGE OF MOTION, WEAR AND IMPINGEMENTS)



Abstract

Hip-spine relationships should be better investigated in THP as lumbo-sacral orientation in the sagittal plane plays a critical role in the function of the hip joints. Lateral X-rays showing spine and hips together in standing, sitting or squatting positions characterize the adaptations of the sagittal balance and the functionnal interactions between hips and spine.

Acetabular cup implantation has to be planned for frontal inclination, axial anteversion, and sagittal orientation. The later refers to the sacro-acetabular angle, key-point in the spine – hip relationships, and that is redefined by the surgeon at the time of implantation.

Usual standard CT-sections are biased for evaluating acetabular anteversion. The conventional CT procedure does not refer to the pelvic bony frame and. the measured anteversion is a projected angle on a transverse plane, depending on the pelvic adaptation in lying position. This measured angle is often considered as anatomical anteversion, leading to some confusion. Therefore this angle is only a “functional” supine anteversion, reflecting the anterior opening angle of the acetabulum in a specific position. According to the sagittal orientation of the pelvis, the true functional acetabular orientation can virtually be assessed in various postures from adjusted CT-scan sections.

The EOS low irradiation 2D-3D X-ray scanner is an innovative technology already used for global evaluation of the spine. This technology allows simultaneously “full body” frontal and lateral X-rays with the patient in standing, sitting or squatting positions; a tridimensionnal patient specific bone recontruction can be performed and the cup anteversion can be directly assessed according to the position.

We investigated the lumbo-pelvic parameters influencing the tridimensionnal orientation of the acetabulum. We compared the data obtained for real postural situations using the EOS system and the measures from plane X Rays and classical CT scan cuts replicating standing, and sitting positions.368 patients with cementless THP were involved in a prospective follow-up protocol. Sacral slope and pelvic tilt, incidence angle, acetabular frontal and sagittal inclination were evaluated on AP and lateral standard XRays. Functionnal anteversion of the cup has been measured using a previously described protocol with CTscan cuts oriented according to standing and sitting sacral slope. The mean difference between CTscan and EOS system was 4,4° with comparable accuracy and reproductibility.

Sacral slope decrease in sitting position was linked to anteversion increase (38,8° SD 5,4°). Sacral slope increase in standing position was linked to lower ante-version (31,7° SD 5,6°). The anatomical acetabular anteversion, the frontal inclination, and the sagittal inclination were functional parameter which significantly varied between the standing, sitting, and lying positions. We noticed that the acetabular parameters in lying position highly correlated to the one in standing position, while poorly correlated with the one in sitting position. The difference between the lying and the sitting positions was about 10°, 25°, and 15° for the cup anteversion (CA) and the frontal and sagittal inclinations (FI,SI) respectively. The poor correlation between the lying and sitting positions suggests that the usual CT scan protocol is biased and not fully appropriate for investigating the cases of posterior THP dislocation and subluxation, which happen in sitting position. On the contrary, a strong correlation was observed between lying and standing measurements with all the acetabular parameters (CA,FI, SI), suggesting that the classical CT assessment of the cup anteversion remains an interesting source of information in case of anterior THP

Each patient is characterized by a morphological parameter, the incidence angle. High incidence angle is linked to low acetabular anteversion, increasing the instability risk and anterior impingement in sitting and squatting position; higher anteversion angles are observed in low incidence angle patients, leading to more internal rotation of the hip in any position.

Lumbo-sacral orientation in the sagittal plane influences the tridimensionnal orientation of the acetabulum, especially for anteversion. Aging of the hip-spine complex is linked to progressive pelvic posterior extension. Impingement phenomenons, orientation of stripe wear zones and some instability situations can be interpreted according to those data.

This study points out the opportunity to adjust the CT scan sections to the sacral slope in functional position for properly investigating the orientation of the acetabular cup, mainly in case of posterior dislocation.

In addition, the mobility of the lumbo-sacral junction could be a crucial parameter in the mechanical functioning and the stability of a THP due to its impact on sacral slope and pelvic tilt. Therefore we also recommend doing dynamic lateral radiographs of the lumbo sacral junction in standing and sitting position for planning a THP implantation in order to detect stiff lumbosacral junction or sagittal pelvic malposition.