header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

FAILURE OF OSTEOINTEGRATION OF A PROXIMALLY-COATED NON-CEMENTED TAPERED STEM: IS THERE AN ANATOMIC PREDISPOSITION?



Abstract

Proximally-coated non-cemented tapered femoral stems have demonstrated excellent long-term clinical results. However, there is sparse literature reporting the incidence of failure of osteointegration in patients with this stem design. The aim of this study is to report this incidence and identify factors which may increase its risk.

206 elective primary total hip arthroplasties were performed consecutively with a single stem design over a three-year period. All patients were evaluated clinically and radiographically. Radiographic parameters were analyzed for any potential risk factors that may predispose to failure of osteointegration.

Three of 206 hips failed to osteointegrate and subsequently underwent revision surgery, for an incidence of 1.5%. The average time to revision was 1.2 years. The presenting complaint was persistent pain and radiographs revealed a progressive linear lucency at the proximal implant-bone interface in all three patients. Each patient had been implanted with a large-sized stem that had achieved a diaphyseal fit radiographically. This cohort had a statistically lower canal-flare index (p < 0.05) when compared to the rest of the study group. At the time of surgery, all stems were found to be loose and were easily removed.

Failure of osteointegration in this type of stem is an uncommon but serious complication that may necessitate revision surgery. Risk factors predisposing to a failure to osteointegrate are a mismatch between the patient’s proximal femoral geometry and the stem, specifically a large stem in a Dorr type C femur, leading to a diaphyseal rather than a metaphyseal wedge.

Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net