header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

APPROPRIATE COMBINED ANTEVERSION FOR PATIENTS WITH EXCESSIVE ANTEVERTED OR RETROVERTED FEMURS IN TOTAL HIP ARTHLOPLASTY AND THE UTILITY OF THE CHANGEABLE NECK



Abstract

Combined anteversion (CA) is defined as the sum of the anteversions of acetabular and femoral components. In this study, we determined the appropriate CA in a variety of femoral versions using a total hip arthroplasty model. In addition, we also examined the usefulness of a changeable neck to improve range of hip motion in these cases.

Using a THA model, the range of motion (ROM) was tested in various CA values obtained by changing the anteversion of a cup in six increments after setting the femoral anteversion to 20° or 60° anteversion and 20° retroversion. The angle of the changeable neck was changed in 11 increments of 5°. To evaluate stability, the range of internal rotation at 90° flexion, the external rotation at 0° extension, and the range flexion was measured when any impingement occurred prior to dislocation. We defined the required ROM that met 40° internal rotation, 30° external rotation, and 110° flexion.

In normal 20° anteversion group, the required ROM was achieved with CA between 30° and 50° without using any changeable necks. In excessive anteversion 60° group, the range of external rotation was less than 10° even when the acetabular component was set 10° retroverted, because of the bone impingement between the greater trochanter and the posterior acetabulum. When 25° retroverted changeable neck was used, ROM improved to 30° external rotation and satisfied the required ROM. In 20° retroversion group, the internal rotation angle was 31° even when the acetabular component was opened 35° anteverted, because of anterior neck-liner impingement. When 25° anteverted changeable neck was used, ROM improved to 39° internal rotation and 130° flexion.

In cases with normal anteversions, the required ROM can be achived by adjusting CA. In cases with excessive anteversion or retroversion, there was a limitation of the CA adjustment. The use of changeable necks allows for further improvement of ROM by compensating femoral anteversions.

Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net