header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

MID-TERM PERFORMANCE OF THE THIRD GENERATION CERAMIC ON CERAMIC ARTHROPLASTY



Abstract

The purpose of this study is to analyze clinical and radiological results of total hip arthroplasty using the 3rd generation ceramic on ceramic articular surface.

Between July 1999 and May 2005, 339 hips of 250 patients had primary cementless total hip arthroplasty with the 3rd generation ceramic on ceramic bearing implants. And 325 hips of 236 patients were followed up over 3 years. Male were 168 patients(237 hips) and female were 68 patients(88 hips). The mean age at the time of operation was 47.3(range, 25~76) years old and the mean follow up period was 62.4(range, 36~107.6) months. The preoperative diagnoses were osteonecrosis of the femoral head (ONFH) in 250 hips, secondary osteoarthritis in 55 hips(dysplasia in 35, infection sequalae in 12, LCP in 2, CDH in 2), hemophilic arthropathy in 9 hips, ankylosing spondylitis in 7 hips etc.

We used Bicontact system(Aesculap, Germany) in 65 hips, Secur-FitTM(Stryker Howmedica Osteonics, USA) in 206 hips, Trilogy ABTM (Zimmer, USA) in 54 hips. Clinically, Harris Hip Score, thigh pain, squeaking and other complications were evaluated. Radiologically, the serial radiographs were analyzed.

Clinically, the Harris hip score was improved from preoperative 66.0(19~91) to 96.2(58~100) at the last follow-up. Radiologically, there was no loosening of implants and visible wear and osteolysis. Heterotopic ossifications were noted in 5 cases. In complications, there was dislocation in one case, periprosthetic fracture in 2 cases and thigh pain in 9 cases. Intermittent squeaking sound has occurred in 8 cases(2.5%). Among these, one case of loud squeaking which happened after fall down had revision surgery. There was no infection and fracture of ceramic implant.

Our midterm results of THA with the 3rd generation ceramic bearing system were very satisfactory and demonstrated that the 3rd generation ceramic bearings remain as an excellent bearing choice because of their superior wear characteristics. However, the results of this study suggests that the squeaking would be one of strong potential risk factors for failure of ceramic on ceramic total hip arthroplasty and we must be very cautious to prevent squeaking.

Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net