header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

IN VIVO STUDY OF SQUEAKING FREQUENCIES IN CERAMIC-ON-CERAMIC THR. COMPARISON WITH IN VITRO VALUES



Abstract

Introduction: The goal of the study was to compare the squeaking frequencies of Ceramic-on-Ceramic THR in-vitro and in-vivo among patients who underwent THR.

Method: Four patients, who underwent THR with a Ceramic-on-Ceramic THR (Trident®, Stryker®) presented a squeaking noise. The noise was recorded and analysed with acoustic software (FMaster®). In-vitro 2 alumina ceramic (Biolox Forte Ceramtec®) 32 mm diameter (Ceramconcept®) components were tested using a PROSIM® hip friction simulator. The cup was positioned with a 70° abduction angle in order to achieve edge loading conditions and the head was articulated ± 10° at 1 Hz with a load of 2.5kN for a duration of 300 cycles. Tests were conducted under lubricated conditions with 25% bovine serum and with the addition of a 3rd body alumina ceramic particle (200 μm thickness and 2 mm length).

Results: In-vivo, recordings had a dominant frequency ranging between 2.2 and 2.4 kHz. In-vitro no squeaking occurred under edge loading conditions. However, with the addition of an alumina ceramic 3rd body particle in the contact region squeaking was obtained at the beginning of the tests and stopped after ~20 seconds (dominant frequency 2.6 kHz).

Discussion and Conclusion: Squeaking noises of a similar frequency were recorded in-vitro and in-vivo. In-vitro noises followed edge loading and 3rd body particles and despite, the severe conditions, squeaking was intermittent and difficult to reproduce. The lower frequency of squeaking recorded in-vivo, demonstrates a potential damping effect of the soft tissues. No damage was observed on the components, however, the test duration was very short. Squeaking may be related to third body particles that could be generated by wear or impingement between the femoral neck and the metal back. Cup design seems to be of particular importance in noisy hip, leading to a high variability of squeaking rate according to the implants.

Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net