header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

HOW DO LIGAMENT CONSTRAINTS INFLUENCE THE WEAR BEHAVIOUR OF KNEE IMPLANTS IN A KNEE SIMULATOR?



Abstract

For wear testing of knee implants, ISO 14243 is the most used testing protocol. In force control, this standard requires linear motion restraints for simulation of ligaments. The aim of this study was to investigate if a nonlinear, physiological motion restraint would influence the wear behaviour of the implants.

A wear study was performed on a highly conforming knee implant design. Three implants were tested forced controlled according to ISO 14243-1 on an AMTI knee simulator. Linear motions restrain of 30 N/mm for AP-translation and 0.6 Nm/° for IE-rotation were applied as required per ISO 14243-1. A second wear test was performed on the same implant design. Based on the data given by Kanamori et al. and Fukubayashi et al., a physiological, nonlinear ligament constraint model (sectioned ACL) was adopted and implemented in the simulation. The implants were pre-soaked and a soak controls was used. Wear was measured gravimetrically.

A mean gravimetric wear rate of 2.85 mg/10E6 cycles was found for the implants which were tested using a linear motion restraint as required per ISO 14243-1. Simulating a physiological, nonlinear motion restraint resulted in a 60% increase in gravimetric wear (mean gravimetric wear rate: 4.75 mg/10E6 cycles). As expected, the kinematics of the implants differed between wear tests. The mean AP-translation increased from 2.89 mm (linear motion restraint) to 4.82 mm (physiological motion restraint). A similar behaviour was observed for the IE-rotation. The IE-rotation increased from 4.09° (linear motion restraint) to 5.94° (physiological motion restraint).

The reaction of the ligaments is not linear in the human knee joint. This study showed that wear and kinematics change when simulating physiological ligament reactions. Wear increased by 60%, an effect which can likely be credited to fundamental differences in kinematics. The ACL is commonly sacrificed during surgery. Thus, more attention should be paid to ligament simulation when performing wear tests on knee implants.

Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net