SQUEAKING FOLLOWING CERAMIC ON CERAMIC TOTAL HIP ARTHROPLASTY.



Abstract

Audible squeaking following ceramic-on-ceramic total hip arthroplasty (THA) is a rare but troublesome problem. We retrospectively reviewed records of 1002 patients where a ceramic-on-ceramic THA had been done during the study period. Fifteen patients complained of squeaking, at any time following their arthroplasty. Fourteen of these 15 patients were evaluated clinically and radiologically. The demographics of these patients were compared to that of all the other patients who did not have squeaking following ceramic-on-ceramic THA. The radiographic data was compared to a control group matched for age, sex, body mass index (BMI), primary diagnosis, type of implant, date of surgery and length of follow-up.

There were 12 males and 2 females of a mean age of 44.5 years (range, 25–65 years). These 14 patients were found to have significantly higher BMI of 25.98 kg/m2 (range, 21.6–32.3 kg/m2) as compared to the other patients who had ceramic-on-ceramic THA (mean, 23.61 kg/m2; range, 15.8 –30.3 kg/m2) (p=0.005). The lateral opening angle was found to be significantly lower (mean, 34°; range 29°–40°) in these patients than the matched control group (mean, 38°; range 30°–41°) (p=0.016). Mean acetabular anteversion was 22° (range 9°–37°), which was not significantly different to that of the matched controls (mean 23°; range 2°–33°) (p=.957). Limb length shortening of more than 5mm was observed in 12 of the 14 (85.7%) patients as compared to only 4 of 14 (28.6%) patients in the matched control group. Two patients had intermittent squeaking while the other 12 had continuous squeaking. Flexion and sitting cross legged were identified as the movements which most commonly (11 of 12) resulted in squeaking. Mean Harris hip score (HHS) improved from 44 (range, 19–66) to 94 (range, 88–100) and most patients (13 of 14) were satisfied with the outcome of the surgery.

Thus the incidence of squeaking was found to be low (1.5%, 15 of 1002) in our series. We identified high BMI, decreased lateral opening angle and limb length shortening as factors associated with occurrence of squeaking. Proper patient selection, implant placement, and avoidance of limb length discrepancy are likely to further reduce the incidence of this complication of ceramic-on-ceramic THA.

Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net