header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

Upper Limb

Long-term survival of the Souter-Strathclyde total elbow replacement in patients with rheumatoid arthritis



Download PDF

Abstract

Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components.

The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants.


Correspondence should be sent to Dr M. Ikävalko; e-mail: mikko.ikavalko@fimnet.fi

For access options please click here