ACCURATE SCALING OF PELVIC DIGITAL X-RAYS WITHOUT A “CALIBRATION OBJECT”



Abstract

Most hospitals have introduced digital radiography (PACS) systems. Accurate pre-operative templating prior to hip arthroplasty requires precise information on the magnification of the digital image. Without this information the benefits of expensive digital templating programs (Orthoview-£10000) cannot be realised.

To determine the magnification of a digital image involves the placement of a “calibration object” at the level of the hip joint. This is unpopular with patients and radiographers alike. We describe a method that requires a single measurement to be made from the greater trochanter to the digital film.

An AP pelvis x-ray was taken of 50 patients with hip replacements. The “predicted” magnification was calculated using the new method. As the size of the head of the prosthesis was known the “actual” magnification could be calculated also. There was no significant difference at 0.05, Wilcoxon T, 2-tail test.

Conventional radiography, which assumes a magnification of 20%, results in errors up to 11%. Templating may therefore predict an incorrectly sized prosthesis. Our method is as accurate as methods using a calibration object whilst being acceptable to patients and staff. Its use should lead to more accurate pre-operative templating prior to total hip arthroplasty

Correspondence should be addressed to Major M Butler RAMC, 44 Theynes Croft, Long Ashton, Bristol, BS41 9NA, England.