THE EFFECT OF HIP ROTATION ON BONE MINERAL DENSITY (BMD) AFTER RESURFACING TOTAL HIP ARTHROPLASTY (RTHA)



Abstract

Resurfacing THA is claimed to transfer stress naturally to the femur neck and preserve proximal femoral bone mass postoperatively. DXA is an established method in estimating BMD around a standard THA, but due to the anteversion of the femur neck, rotation could affect the size of the neck-regions and thereby the BMD measurements around a RTHA. To our knowledge, this is the first study to analyze the effects of hip rotation on BMD in the femoral neck around a RTHA.

We scanned the femoral neck of 15 patients twice in each position of 15° inward, 0° and 15° outward rotation, and analyzed BMD in a single and a six-region model. CVs were calculated for BMD in the same position as well as between different positions.

For double measurements in the same position we found mean CVs of 3.1% (range 2.5% – 3.7%) and 4.6% (range 2.2% – 8.6%) in the one- and six-region models, respectively. When the 15° outward position was excluded, the CVs decreased to 2.8% and 4.0%. With rotation, the mean CVs rose to 5.4% (range 3.2%–7.2%) and 11.8% (range 2.7% – 36.3%). This effect was most pronounced in the 6-region model, predominantly in the lateral and distal parts of the femoral neck, where the change was significantly different from the fixated position. For the single-region model 15° rotation could be allowed without compromising the precision.

We conclude that rotation adversely affects the precision of BMD measurements around a RTHA, but in the single-region model smaller rotations can be allowed.

With the hip fixated the six-region model produces low CVs, acceptable for longitudinal studies. For maximal topographical detail we prefer the six-region model and recommend that future longitudinal DXA studies, including RTHA, be performed standardised, Preferably, with the hip in the neutral or internal rotation.

Correspondence should be addressed to EORS Secretariat Mag. Gerlinde M. Jahn, c/o Vienna Medical Academy, Alserstrasse 4, 1090 Vienna, Austria. Fax: +43-1-4078274. Email: eors@medacad.org