MID TERM RESULTS OF THE MORSCHER PRESS FIT ACETABULAR CUP



Abstract

The Morscher press fit acetabular component is a monobloc cup with the polyethylene bonded directly to a titanium mesh shell. There is little published data on the longevity of the Morscher cup apart from the designer’s series. It has been quite widely used in New Zealand since its introduction in 1993. The aim of this retrospective study was to provide an independent mid-term audit of the results of this cup in the New Zealand population.

A retrospective review was undertaken of all Morscher cups implanted at Dunedin Public Hospital or Mercy Hospital by 5 orthopaedic surgeons between 1994 and 1998 with a minimum follow-up of seven years. Clinical and radiological survey was performed with standardised scoring systems. A new method for measuring linear cup wear was developed due to the unusual geometry of the Morscher cup.

136 hips were replaced in 121 patients during the study period 101 were performed in private and 35 in public. Our follow up ranged from 7 to 11 years (mean 8.69). There were 73 males (85 hips) and 48 females (51 hips). The average age of the patients was 57.5yrs (SO 24.97). Pre operatively the mean Merle d’Aubigne score was 9.4 (SO 4.02) and post operatively it was 17.4 (SO 1.58 (p< 0.0001).

There were 2 early revisions: one at 3 days for instability post-operatively and one for an unrecognised intra-operative femoral fracture. One hip required a two stage revision for deep infection at 3 years. Two hips have been revised for polyethylene wear and osteolysis at 7 yrs and 9 yrs and one hip required bone grafting of a large acetabular osteolytic lesion with retention of the cup. A further 4 patients have significant osteolytic defects and 2 have minor osteolytic lesions.

Wear measurements in the unrevised hips have shown a mean linear wear rate of 0.079mm/yr (range 0.000 to 0.222mm). 33% of these hips have a wear rate of > 0.1 mm/yr and 25% have a wear rate of < 0.05mm/yr.

We have found excellent clinical results with the use of the Morscher cup in this relatively young and active cohort. A small group of patients, however, have shown higher wear rates or osteolysis, and we emphasise the importance of continuing radiological review.

Correspondence should be addressed to Associate Professor N. Susan Stott at Orthopaedic Department, Starship Children’s Hospital, Private Bag 92024, Auckland, New Zealand