header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

NEW TYPES OF POLYETHYLENE IN TOTAL HIP REPLACEMENT: PROSPECTIVE AND COMPARATIVE CLINICAL STUDY ON WEAR



Abstract

Introduction and purpose: Highly cross-linked polyethylenes (HXLPE) sterilized in an air-free environment have been used to prevent osteolysis and loosening of implants. This prospective randomized study analyzes the results of a series in which one single type of prosthesis but made with one of two different kinds of polyethylene (PE) has been used.

Materials and methods: We assessed 45 Allofit cups with a Sulene-PE liner (sterilized in nitrogen) and 45 with Durasul-PE liner (highly cross-linked) associated with an Alloclassic stem (femoral head: 28 mm) with a minimum follow-up of 5 years (mean 66.3 months). Femoral head penetration was analyzed by means of a digital program at 6 weeks, and at 6 and 12 months and once a year, using the Dorr method given the non-spherical shape of the cups.

Results: All assessed hips had good outcomes determined clinically and by x-rays. There was no loosening of any component. There were no radiolucent lines or osteolysis. Femoral head penetration at 6 weeks was 47.4% less in the Durasul group (0.19+0.06 mm for the Sulene-PE and 0.09+0.03 for the Durasul-PE [p< 0.0001]). Mean annual penetration was 20% less in the Durasul group (0.04+0.02 and 0.008+0.008 [p< 0.0001] respectively. The differences increased by the third year. Mean penetration at 5 years was 39.1% less in the Durasul group (p< 0.0001).

Conclusions: There was significant femoral reduction in the Durasul-PE group. Long term results are necessary to confirm that these prostheses lead to a lower rate of osteolysis.

The abstracts were prepared by E. Carlos Rodríguez-Merchán, Editor-in-Chief of the Spanish Journal of Orthopaedic Surgery and Traumatology (Revista de Ortopedia y Traumatología). Correspondence should be addressed to him at: Sociedad Española de Cirugía Ortopédica y Traumatología, calle Fernández de los Ríos 108, 28015-Madrid, Spain