The effect of the offset humeral head on the micromovement of pegged glenoid components

A COMPARATIVE STUDY USING RADIOSTEREOMETRIC ANALYSIS



Download PDF

Abstract

In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with offset humeral head components. Over a period of 24 months the relative movement of the glenoid component with respect to the scapula was measured using radiostereometric analysis.

Nine glenoids needed reaming for erosion. There was a significant increase in rotation about all three axes with time (p < 0.001), the largest occurring about the longitudinal axis (anteversion-retroversion), with mean values of 3.8° and 1.9° for the non-offset and offset humeral head eroded subgroups, respectively. There was also a significant difference in rotation about the anteversion-retroversion axis (p = 0.01) and the varus-valgus (p < 0.001) z-axis between the two groups. The offset humeral head group reached a plateau at early follow-up with rotation about the z-axis, whereas the mean of the non-offset humeral head group at 24 months was three times greater than that of the offset group accounting for the highly significant difference between them.


Correspondence should be sent to Dr D. Nuttall; e-mail: uplimb_dn@hotmail.com

For access options please click here