header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

COMPUTER-ASSISTED HIP RESURFACING SURGERY USING THE ACROBOT NAVIGATION SYSTEM



Abstract

Last year at CAOS UK we reported on the development of the Acrobot® Navigation System for accurate computer-assisted hip resurfacing surgery. This paper describes the findings of using the system in the clinical setting and includes the improvements that have been made to expedite the procedure. The aim of our system is to allow accurate planning of the surgery and precise placement of the prosthesis in accordance with the plan, with a zero intra-operative time penalty in comparison to the standard non-navigated technique.

The system uses a pre-operative CT-based plan to allow the surgeon to have full 3D knowledge of the patient’s anatomy and complete control over the sizes and positions of the components prior to surgery.

At present the navigation system is undergoing final clinical evaluation prior to a clinical study designed to demonstrate the accuracy of outcome compared with the conventional technique. Whilst full results are not yet available, this paper describes the techniques that are being used to evaluate accuracy by comparing pre-operative CT-based plans with post-operative CT scans, and gives initial results.

This approach provides a true measure of procedure outcome by measuring what was achieved against what was planned in 3D. The measure includes all the sources of error present within the procedure protocol, therefore these results represent the first time that the outcome of a navigated orthopaedic procedure has been measured accurately.

Correspondence should be addressed to Mr K. Deep, General Secretary CAOS UK, 82 Windmill Road, Gillingham, Kent ME7 5NX UK. E Mail: caosuk@gmail.com