FUNCTIONAL OUTCOME OF THE ROBOT ASSISTED MINIMAL INVASIVE UNICOMPARTMENTAL KNEE ARTHROPLASTY AT 18 MONTHS



Abstract

A Prospective, randomised controlled trial demonstrates superior outcomes using an active constraint robot compared with conventional surgical technique in unicompartmental knee arthroplasty (UKA). Computer assistance should extinguish outliers in arthroplasty, with robotic systems being able to execute the preoperative plan with millimetre precision.

We used the Acrobot system to deliver tailor made surgery for each individual patient. A total of 27 patients (28 knees) awaiting unicompartmental knee arthroplasty were randomly assigned to have the operation performed either with the assistance of the Acrobot or conventionally. CT scans were obtained with coarse slices through hips and ankles and fine slices through the knee joint. Preoperative 3D plans were made and transferred to the Acrobot system in theatre, or printed out as a conventional surgical aid. Accurate co-registration was confirmed, prior to the surfaces of the femur and tibia being milled. The outcome parameters included measurements of the American Knee Society (AKS) score and Western Ontario and McMaster Universities Osteoarthritis (WOMAC) index. These measurements were performed pre-operatively and at six, 18 weeks, and 18 months post-operatively. After 18 months two UKA out of the conventional trial (n =15) had been revised into a total knee replacement (TKA), whereas there were no revisions in the Acrobot trial group (n = 13).

Using an active constrained robot to assist the surgeon was significantly more accurate than the conventional surgical technique. This study has shown a direct correlation between accuracy and improvement in knee scores at 6, 18 weeks and 18 months after surgery. At 18 months there continues to be a significant improvement in the knee scores with again a marked correlation between radiological accuracy and clinical outcome with higher accuracy leading to better function based on the WOMAC and American Knee Society Score.

Correspondence should be addressed to Mr K. Deep, General Secretary CAOS UK, 82 Windmill Road, Gillingham, Kent ME7 5NX UK. E Mail: caosuk@gmail.com