header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

P57 THE CREEP BEHAVIOUR OF LEADING CEMENT BRANDS



Abstract

The visco-elastic behaviour of cement, is a key feature of cement-implant performance in total hip arthroplasty.

The aim of this study was to describe the creep behaviour of the leading plain bone cements under standardised physiological in-vitro conditions.

Methods: Cements were mixed under vacuum conditions as per manufacturers instructions. Moulds were used to to produce beams of standard dimensions. These were stored in saline at 37oC for 21 days to ensure thorough polymerisation. Under the same conditions, the beams were tested for 72 hours in a 12-station quasi-static creep rig, using a four-point bending configuration. The rig applied a constant stress of 8MPa to each beam and the deflection was recorded at 8-minute intervals by a data-logging device. The porosity was measured in the mid-cross section of each beam sample using a digital image technique.

The cements tested were Palacos R, CMW1 and Smartset GHV and Surgical Simplex P.

All data were analysed using ANOVA with Bonfer-roni post-hoc test (SPSS).

Results: Palacos R exhibited the highest mean deflection at 72 hours (0.86+/- 0.21mm) followed by Surgical Simplex P (0.85 +/- 0.18mm), CMW1 (0.72 +/- 0.09mm) and Smartset GHV (0.60 +/- 0.16mm). The difference between the two DePuy cements and Palacos R (p=0.03) and Surgical Simplex P (p=0.04) were statistically sig-nificant. None of the beams failed during the test. The creep behaviour correlated with the cross-sectional porosity measurements.

Conclusions: This study has shown that there are sig-nificant differences in the creep bahaviour of the leading medium and high viscosity bone cements. In particular Palacos R and Surgical Simplex P demonstrate ‘High’ creep and the DePuy cements ‘Low’ creep. Creep appears sensitive to subtle changes in the composition of the material. This may be reflected in the clinical behaviour of different bone cements and stresses the importance of the time-dependent properties of PMMA.

Correspondence should be addressed to Mr Carlos Wigderowitz, Senior Lecturer, University Department of Orthopaedic and Trauma Surgery, Ninewells Hospital and Medical School, Dundee DD1 9SY.