header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

CONTEMPORARY CAUSES OF PREMATURE FAILURE OF PRIMARY TOTAL KNEE REPLACEMENT.



Abstract

The purpose of this study was to determine the incidence of revision total knee replacement (TKR) within 5 years of the index procedure at a large multi-surgeon unit using a single prosthesis and to determine the cause of failure of those implants.

This was a retrospective review of all primary Kine-max Plus TKR performed at the Avon Orthopaedic Center between 1.1.1990 and 1.1.2000. Cases were identified that required revision arthroplasty in any form within 5 years of the index procedure. Case notes and Xrays were reviewed to determine causes of failure.

There were 2826 primary Kinemax Plus TKR performed during the study period. Of these 20 were known to have required revision surgery within 5 years. 8 were revised for deep infection of the prosthesis and 12 for aseptic causes. The overall incidence of premature failure of the Kinemax Plus TKR at 5 years was 0.71%. The incidence of aseptic premature failure at 5 years was 0.42%. Detailed examination of the clinical records indicated that some form of technical error at the time of the index arthroplasty was responsible for the early failure of 6 prostheses. This equates to 0.21% of the procedures performed. Aseptic loosening of the remaining 6 cases could not be attributed to a specific cause.

Correspondence should be addressed to Mr Tim Wilton, BASK at the Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PE.