THE EXETER UNIVERSAL CEMENTED FEMORAL COMPONENT AT 14 TO 17 YEARS: AN UPDATE ON THE FIRST 325 HIPS.



Abstract

Introduction: We describe an update of our experience with the implantation of the first 325 Exeter Universal hips. The fate of every implant was known.

Methods: This is a review of the prospective data collected from the first 325 Exeter Universal stems (309 patients) inserted between March 1988 and February 1990. The procedures were undertaken by surgeons of widely differing experience. Clinical and radiological review was performed at a mean of 15.4 years.

Results: At last review 185 patients had died (192 hips). 104 hips remain in-situ. Survivorship at 17 years with revision for femoral component aseptic loosening was 100% (95% CI 97 to 100), with revision for acetabular component aseptic loosening was 92.85% (95% CI 87.0 to 96.2) and with any re-operation as the endpoint was 83.99% (95% CI 77.1 to 90.27). 12 patients (12 hips) were not able to attend for review due to infirmity or emigration, and scores were obtained by phone (x-rays were obtained in 4 patients). Mean D’Aubigné and Postel scores (Charnley modification) at review were 5.4 for pain and 4.5 for function. The mean Oxford score was 21.46 +/− 9.52 and the mean Harris score 73.35 +/− 17.32. On radiological review there were no femoral component failures. Three sockets (2.9%) were loose as demonstrated by migration or change in orientation (two patients were asymptomatic) and 5 sockets (4.8%) had radiolucent lines in all 3 zones but no migration. There are two patients awaiting socket revision.

Discussion: The high rate of clinical and radiological success of the Exeter Universal stem seen at 12 years is sustained at 17 years. Further cup failures have occurred but overall survivorship remains good. With the favourable long-term behaviour of the original Exeter stem, we feel optimistic that good function of the Universal stem will continue through the third decade.

Correspondence should be addressed to The Secretary, BHS, c/o BOA, The Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PE.