header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

EXCELLENT MID-TERM RESULTS USING A PRESSFIT POROUS-COATED COBALT-CHROMIUM ALLOY CUP FOR ACETABULAR REVISION.



Abstract

Introduction: This study is a prospective series using a porous-coated cobalt-chromium alloy cup augmented with screw fixation for acetabular revisions.

Methods: Between August 1997 and December 2001, eighty-five consecutive cementless acetabular revisions (81 patients) were performed. The mean age at operation was 64 years. Using the AAOS classification of acetabular defects there was one type I defect, 25 type two defects and 59 type III defects. Eighty-three cases were available for review (98%) with an average follow up of 6 years 6 months (range 3 months – 110 months). Clinical outcome was measured using the Charnley Hip Score and radiological assessment by plain radiographic measurement.

Results: The Charnley Hip Score had improved from a pre-operative average of 7.52 (range 4–11) to the latest score of 14.84 (range 8 – 18). Significant cup migration occurred in only one case, which did not require revision. Sixteen cases demonstrated non-progressive radiolucent lines in one or two Delee and Charnley zones but none extended to all three zones. There were no cases of significant osteolysis. There were five dislocations none requiring re-operation; one revision was carried out for deep infection and one liner exchange at time of stem revision for subsidence. The results of Kaplan-Meier survival analysis using revision for all causes as the endpoint was 98.8% (95% confidence limits 0.964 to 1) at seventy-three months.

Discussion: This press fit porous-coated cobalt-chromium alloy cup augmented with screw fixation for acetabular revision surgery produced excellent midterm results. Changes in cup design and material should only be undertaken with consideration of such results.

Correspondence should be addressed to The Secretary, BHS, c/o BOA, The Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PE.