VALIDATION OF A MODEL TO ESTIMATE THE IN-VIVO PATELLOFEMORAL CONTACT FORCES OF TKR



Abstract

Patellofemoral pain is a significant problem for patients with Total Knee Replacements (TKRs). It is hypothesized that pain is related to high patellofemoral forces (PFF). The aim of this study is to validate a model to estimate PFF after TKR, using a combination of non-invasive measurement and theoretical modeling.

Experiments were performed on four cadaver knee specimens to compare the PFF and the quadriceps force (QF) estimated by a model, with those measured using force transducers. Each knee was tested in its initial state and after implantation of three Scorpio designs: Cruciate Retaining (CR), Posterior Stabilised (PS), and the Posterior Stabilised Mobile Bearing (PS+). Each knee was extended/flexed under a simulated quadriceps load with 3 kg hung from the distal tibia. Relative movement of the bones was measured using a Vicon 612 motion analysis system. A 6DOF force transducer was used to measure PFFs and a uni-axial transducer was used to measure QFs. A fluoroscope simultaneously captured images of the leg extension activity. Parameters measured from the images were used as inputs to the model.

The measured and estimated PFF and QF matched closely between 20o and 80o of knee flexion for the TKRs. At higher flexion angles, the model overestimated the PFF by a maximum of 23N (7.6% max) for the PFF and by 31N for the QF (10.3% max). The estimated and measured Patellar Flexion Angles (PFA) were within 3.5o throughout the flexion range.

The model accurately predicts sagittal plane patellar kinematics and kinetics, using only fluoroscopy and externally measured forces as inputs. However, the model has a limitation in assuming that the extending moment is only due to the quadriceps.