CT-FREE CUP ORIENTATOR USING TILT SENSORS IN THR



Abstract

The purpose of this research is to propose CT-free cup orientator using tilt sensors without expensive point tracking devices in total hip replacement. In the case of using a mechanical guide, the accuracy of cup orientation can be sacrificed because of change of the patient’s posture during procedure. Several navigation systems have been introduced to secure an accurate position and orientation of the implant in THR. These systems are expensive and have some weakness due to possible interference inoptical measurement. Our orientator employs a T-bar shaped gauge and economic tilt sensors to secure a fairly orientation of acetabular cup inTHR.

The T-bar gauge having three feet with adjustable distance is designed to obtain the anatomical landmarks concurrently. Each foot is placed on the anatomical landamark of the sawbone. The gauge has its own tilt sensor to identifiy the tilt angle of the guage using AD input board. Similary, the cup positioning tool and dynamic reference base (DRB) have their own tilt sensors. The experimental procedures of CT-free cup orientator are done as follows:

  1. Place the T-bar gauge in right place on the pelvis by setting three feet on the ASIS and pubic.

  2. Attach DRB to pelvis and align its orientation parallel to the T-bargauge.

  3. Align the tilt sensor of the cup positioner parallel to DRB.

We define errors as difference between experimental data and ground truth obtained by Micro-Scribe (Immersion Inc.) Errors of the cup in abduction and anteversion were 1.2 and 1.0 degrees respectively when the test is performed on a sawbone.

We analyzed the causes of error to improve the accuracy of our cup orientator. Measuring landmarks and aligning three tilt sensors seemed to cause some errors. Base on this study, we expect to make an experiment on cadaver.

Correspondence should be addressed to Richard Komistek, PhD, International Society for Technology in Arthroplasty, PO Box 6564, Auburn, CA 95604, USA. E-mail: ista@pacbell.net

Acknowledgments: This work has been supported by the Human-friendly WelfareRobot System Engineering Research Center (HWRS-ERC) at KAIST in Korea