THE INFLUENCE OF FEMORAL CONDYLAR LIFT OFF ON THE WEAR OF FIXEBEARING ARTIFICIAL KNEE JOINTS



Abstract

Considerable differences in kinematics between different designs of knee prostheses and compared to the natural knee have been seen in vivo. Most noticeably, lift off of the femoral condyles from the tibial insert has been observed in many patients. The aim of this study was to simulate lateral femoral condylar lift off in vitro and to compare the wear of fixed bearing knee prostheses with and without lift off.

Twelve PFC Sigma cruciate retaining fixed bearing knees (DePuy, Leeds, UK) were tested using six station simulators (Prosim, Manchester, UK). The kinematic input conditions were femoral axis loading (maximum 2.6 kN), flexion-extension (0–58°), internal/external rotation (±5°) and anterior/posterior displacement (0–5 mm). Six knees were tested under these standard conditions for 4 million cycles. Six knees were tested under these conditions with the addition of lateral femoral condylar lift off, for 5 million cycles. The lubricant used was 25% newborn calf serum. Wear of the inserts was determined gravimetrically.

Under the standard kinematic conditions the mean wear rate with 95% confidence limits was 8.8 ± 4.8 mm 3/million cycles. When femoral condylar lift off was simulated the mean wear rate increased to 16.4 ± 2.9mm 3/million cycles, which was statistically significantly higher (p < 0.01, Students t-test). The wear patterns on the femoral articulating surface of all the inserts showed more burnishing wear on the medial condyle than the lateral. However, in the simulation of lift off the medial condyle was more aggressively worn with evidence of adhesion and surface defects.

The presence of lateral femoral condylar lift off accelerated the wear of PFC Sigma cruciate retaining fixed bearing knees. The lateral lift off produced uneven loading of the bearing, resulting in elevated contact stresses and hence more wear damage to the medial side of the insert. The implications of condylar lift off include increased wear of the polyethylene and possible osteolysis.