DETERMINATION OF FEMORAL COMPONENT SIZE IN UNICOMPARTMENTAL KNEE REPLACEMENT



Abstract

The purpose of the study: to determine if Patient height-and gender could be used to predict component size With a minimally invasiveapproach for unicompartmental knee replacement.

Material and methods: One hundred x-rays of patients (44 men, 56 women), who had undergone Oxford UKR, were reviewed. The preoperative radiographs were assessed for component size using the standard template. The postoperative x-rays were reviewed to determine-whether the ideal component size had been used or if not what could be the most appropriate. Patient’s height was recorded. The proportion of patients for whom an appropriate size could be selected by either template or height measurements was calculated.

Current templating system accurately predicted the ideal size in 67%. In no case was the size incorrect by more than one size. The following size bands were set according to height. For men: size small in patients less than160 cm, medium less than 170 cm and large less than 180cm. For women: size small in patients less than 165 cm, medium less than 175cm and large less than 185 cm. Height accurately predicted the ideal size in75%. In no case was the assessment of component size incorrect by more than one size.

As the Oxford femoral component is spherical, its size is not critical and it is acceptable to use one size too large or too small. Both height and templating safely predicted an acceptable size in all cases and predicted the ideal size in about 70% > Conclusion: Gender specific height should be used to predict the component size in situations were templating is difficult as in digital x-rays orsuperimposition of the two femoral condyles, and non-standardised x-raymagnification.

Correspondence should be addressed to Richard Komistek, PhD, International Society for Technology in Arthroplasty, PO Box 6564, Auburn, CA 95604, USA. E-mail: ista@pacbell.net