header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

INTRAMEDULLARY GUIDE FOR FEMORAL COMPONENT POSITIONING IN TKA: HOW ENTRY POINT AFFECTS FEMORAL COMPONENT POSITION



Abstract

Purpose: This cadaveric study examines how changes in femoral entry point for intramedullary instrumentation of total knee replacements affects femoral component positioning.

Methods: Twelve cadaveric lower limb specimens with intact hip, knee and ankle joints were obtained. Total knee navigation instrumentation was secured. Anatomical landmarks required for axes generation were obtained. An initial entry point was made at the center of the distal femur. An intramedullary rod was the introduced into the femur. Five and seven degree cutting blocks were placed onto the rod and positioned against the distal femur with the rotation parallel to the epicondylar axis. The navigation system was then used to generate a varus/valgus angle and flexion/extension angle with respect to the previously generated femoral mechanical axes. This allowed determination of an angle at which the distal femoral cutting block would need to be set to make a neutral distal femoral cut. The guide rod was removed and reinserted five times and measurements recalculated. Data was then collected with entry points 5mm medial, 5mm anterior and 5mm medial and anterior to the initial entry point.

Results: There was no significant difference in varus/ valgus angle with a central compared with 5mm anterior entry point and no difference with a 5mm medial versus 5 mm medial and anterior entry point. The valgus angle required to give a neutral distal femoral cut with a central entry point was 4.98o (SD 0.91o; range 3.5o–6.0o). The valgus angle for a 5mm medial entry point was 6.92o (SD 0.97o, range 5.5o–8.0o). With regards to the sagittal plane a 5mm anterior translation of the entry point changed the flexion/extension angle by 1.58o (SD 0.52o, range 0.5o–2.5o).

Conclusions: Small changes in the entry point can significantly affect component alignment. When moving more medial with the entry point a more valgus angle is required for the cutting block. An entry point at the deepest point of the trochlea may be more reproducible than an anteromedial one but requires a valgus cutting block closer to 5 degrees.

Funding : Commerical funding

Funding Parties : Stryker

Correspondence should be addressed to Cynthia Vezina, Communications Manager, COA, 4150-360 Ste. Catherine St. West, Westmount, QC H3Z 2Y5, Canada