AN ANALYSIS OF FRACTURED DISTAL FIXATION REVISION FEMORAL STEMS – THE ROLE OF LACK OF PROXIMAL SUPPORT, STEM SIZE AND EXTENDED TROCHANTERIC OSTEOTOMY



Abstract

Femoral stem fracture occurred in five (2%) of 283 revision hip arthroplasties when extensively coated, distal fixation femoral implants were used. Common features were complex revisions, high body mass index, poor proximal support, smaller stems (< 13.5 mm) and use of an extended trochanteric osteotomy. Use of strut allografts can reduce femoral stem stresses.

The purpose of this study was to analyze the surgical factors associated with fracture of cementless distal fixation revision femoral stems and supplement this analysis with a finite element analysis.

Data from our institutional joint replacement database identified five (2%) fractured femoral stems (three Solution, two Echelon) in a cohort of two hundred and eighty-three cementless distal fixation femoral stems (one hundred and eighty-two Solution, one hundred and one Echelon) followed over two years. Patient, surgical and implant factors were compared between the fractured and not fractured stems. Finite element analysis was performed on two fractured stem cases to better understand the fracture mechanism.

Common clinical features of the five fractured stems were complex revisions, high patient body mass index, poor proximal bone support, smaller stems (< 13.5 mm) and use of an extended trochanteric osteotomy (ETO) (3/5). Fatigue failure was the mode failure in all cases.

Finite element analysis confirmed that the location of highest stem stress was near the location of the stem fracture and that use of a strut graft over an extended trochanteric osteotomy in patients lacking proximal femoral cortical support decreased the stem stresses by 48%.

Use of cementless distal fixation revision femoral stems is accompanied by a 2% risk of stem fracture in the face of proximal femoral support of the stem and/or use of a stem with a diameter less than 13.5 mm. When an extended trochanteric osteotomy is used, the stem fracture rate increased to 4%. Use of a strut allograft in conjunction with the ETO can reduce stem stresses by 48% and the risk of stem fracture.

Funding: Research support from J& J DePuy and Smith & Nephew