CLINICAL ACCURACY OF FLUOROSCOPIC COMPUTER-ASSISTED PEDICLE SCREW PLACEMENT. A PROSPECTIVE CT ANALYSIS.



Abstract

Using post-operative CT analysis the clinical accuracy of computer-assisted fluoroscopy for the placement of thoracic (n=69) and lumbosacral (n=271) pedicle screws was assessed. All screws were placed using the Fluoro-Nav™ system (Medtronic Sofamor Danek, Memphis, TN, USA). Screw position was completely intrapedicular in 86.5%. There were no clinically significant screw misplacements. Pedicle breaches with a potential for neurological injury (> 2 mm; medial) was 0.6%. The overall pedicle screw misplacement rate in this study is less than or comparable to reported misplacement rates using other techniques. The use of computer-assisted fluoroscopy may improve the safety of pedicle screw placement.

The purpose of this prospective study is to evaluate the clinical accuracy of computer-assisted fluoroscopy for the placement of thoracic (T) and lumbosacral (LS) pedicle screws.

The overall thoracic and lumbar pedicle screw misplacement rate in this study is less than or comparable to reported misplacement rates using other techniques.

The use of computer-assisted fluoroscopy may improve the safety of pedicle screw placement.

Postoperative computed tomographs (CT) of three hundred and forty pedicle screws were independently reviewed. All screws were placed using the Fluoro-Nav™ system (Medtronic Sofamor Danek, Memphis, TN, USA). The relative position of the screw to the pedicle was assessed and graded as follows – A- completely in; B – < 2mm breach; C – 2–4mm breach; D – > 4–6mm breach. If an osseous breach occurred, the direction of the breach was further classified.

Overall screw position was graded A in 86.5% (294/340) of screws (91.1 % (24/271) -lumbosacral and 68.1.0% (47/69)-thoracic). Forty-six pedicle breaches occurred (24 medial; 22 lateral). Thirty-five percent (16/46) of breaches were unavoidable secondary to a pedicle screw that was larger than the size of pedicle (thoracic-13). Pedicle breaches were Grade B in 11.8%, Grade C in 1.5% and Grade D in 0.3% of screws. There were no clinically significant screw misplacements. Pedicle breaches with a potential for neurological (> 2 mm; medial) or vascular injury was 0.6% and 0% respectively.

FluoroNav™ appears to be a safe and practical adjunct for the placement of thoracic and lumbosacral pedicle screws.

Funding: Medtronic-Sofamor Danek – research support