Full Access

SERUM METAL ION LEVELS AFTER METAL-ON-METAL HIP ARTHROPLASTY: DOES BEARING SIZE MATTER?



Abstract

Metal-on-metal (MOM) bearings after total hip arthroplasty are known to elevate the serum concentrations of metal ions, raising concerns about the long-term effects. One potential modifier of ion release is the bearing diameter used. Resurfacing MOM bearings have a large surface area available for corrosion compared to the typical bearing size of 28 mm in total hip replacement (THR) but may benefit from improved lubrication and reduced production of corrodible wear debris. The net effect of these two variables on metal ion release is unknown. In this study, we compared the level of ion release in patients after large bearing MOM hip resurfacing arthroplasty with patient after small bearing MOM THR.

We measured the serum cobalt and chromium levels from 22 patients with large bearing diameter MOM hip resurfacing arthroplasty (Cormet 2000 and Birmingham Hip Resurfacing) and compared them to the serum cobalt and chromium levels of 22 patients with small bearing diameter (28 mm) MOM THR (Ultima). Patients were prospectively matched for activity level, body mass and date after surgery at blood sampling. All were at least 6 months after surgery.

We found the median cobalt and chromium levels after hip resurfacing arthroplasty to be 7.6 times normal (median 38 nmol/L, range 14 to 144 nmol/L) and 10.5 times normal (median 53 nmol/L, range 25 to 165 nmol/ L) respectively. This is compared to 4.4 times normal (median 22 nmol/L, range 15 to 87 nmol/L) for cobalt and 3.8 times normal (median 19 nmol/L, range 2 to 58 nmol/L) for chromium after 28 mm MOM THR (p=0.0021 and p< 0.0001).

Conclusion: Large diameter MOM bearings result in greater release of cobalt and chromium ions than do small diameter MOM bearings. This may be of relevance when the potential effects of long-term exposure to elevated these metal ions is considered.

The abstracts were prepared by Mr Tim Briggs. (Editoral Secretary 2003/4) Correspondence should be addressed to him at Lane Farm, Chapel Lane, Totternhoe, Dunstable, Bedfordshire LU6 2BZ, United Kingdom