TEN YEAR SURVIVORSHIP OF THE ORIGINAL M E MULLER STRAIGHT STEM TOTAL HIP REPLACEMENT SYSTEM



Abstract

To determine the ten-year survivorship of the Original M E Muller Straight Stem Total Hip Replacement System with emphasis on the longevity of the femoral component in accordance with guidelines published by the National Institute of Clinical Excellence, 266 consecutive hip arthroplasties using the above prosthesis were performed by the senior author between 1983 and 1992. 24 patients were lost to follow-up. Of the remaining 242 patients 80 were male and 162 female. The mean age was 67.49 years. The diagnosis for the majority of patients was osteoarthritis of the hip joint. Pre-operative planning was carried out and the patients were scored using Charnley’s modification of the d’ Aubigne and Postel numerical grading system A mono-bloc stem with a 32-millimeter head used via the trans gluteal approach recommended by Muller. Following discharge serial follow-up consisted of both clinical and radiological evaluation. The data was prospectively stored on a Microsoft access database. The survival of the prosthesis using revision for aseptic loosening as an end-point was calculated by actuarial analysis.

135 patients attended their ten-year follow-up. 97% of patients had good to excellent pain relief and improvement in movement of the joint following surgery. 38% had good to excellent mobility with the remaining having restricted mobility due to associated co-morbid factors. Only in 3% of patients was mobility restricted as a result of the arthroplasty. 7 revisions were carried out for aseptic loosening, all as a result of failure of the acetabular component. The cumulative survival for this hip replacement system was 95.9% and that for the femoral component was 100% at 10 years.

The Muller Straight Stem femoral component is based on a press-fit concept and gives predictable long-term results when recommended surgical technique is followed. This series confirms the reliability of the stem design and satisfies the NICE guidelines.

The abstracts were prepared by Mr Tim Briggs. (Editoral Secretary 2003/4) Correspondence should be addressed to him at Lane Farm, Chapel Lane, Totternhoe, Dunstable, Bedfordshire LU6 2BZ, United Kingdom