THE USE OF THE LONG GAMMA NAIL IN PROXIMAL FEMORAL FRACTURES



Abstract

We report the results of the use of the Long Gamma Nail in the treatment of complex proximal femoral fractures in our hospital.

All patients at one hospital treated with the Long Gamma Nail were reviewed. Information collected included the age, sex, type of injury, fracture classification, intra-operative complications, post-operative complications, and survival of the implant and patient.

One hundred nails were reviewed which were inserted in 97 patients. 70 patients were followed up for 1 month or more and their mean follow up was 8 months (range 3 months to 6 years). The mean age was 74 (range 16–98). Twenty were inserted into femurs with metastatic malignancy and four patients were victims of poly-trauma. The average length of the operation was 2 hours 22 minutes. Blood transfusion was required in 74% and on average was 2.5 units. There were 7 significant complications. Five patients underwent revision, 2 to Total Hip Arthroplasty after proximal screw migration and 2 patients required exchange nailing. There was one broken nail and two peri-prosthetic fractures at the tip of the nail.

Success was defined as achievement of stability of fracture until union or death; this was achieved in 15% of cases. The mortality was 7% at 30 days and 17% at one year. One death was directly related to the nail and the rest due to medical co-morbidities. Complication rate fell with increasing experience in the unit. The training of surgeons had no detrimental effect on outcome.

Complex proximal femoral fractures including pathological lesions, subtrochanteric fractures and pertrochanteric fractures with subtrochanteric extensions are difficult to treat, with all implants having high failure rates. The long gamma nail allows early weight bearing and seems effective in treating these difficult fractures. Furthermore the majority of these unstable fractures tend to occur in the very elderly with osteoporosis and other medical co-morbidity. Care should be taken to avoid malpositioning of the implant, as this was the major cause of failure and revision. The length of time surgery may take and the anticipated blood loss should not be underestimated especially when dealing with challenging fractures in frail and elderly patients or those with medical co-morbidity.

The abstracts were prepared by Mr Tim Briggs. (Editoral Secretary 2003/4) Correspondence should be addressed to him at Lane Farm, Chapel Lane, Totternhoe, Dunstable, Bedfordshire LU6 2BZ, United Kingdom