BIHAPRO CUP. MULTICENTRIC STUDY.

7th Congress of the European Federation of National Associations of Orthopaedics and Traumatology, Lisbon - 4-7 June, 2005



Abstract

Introduction.- Some of the recommended alternatives to increase the cementless acetabular cups stability are the plasma sprayed porous coated and HA and the press-fit impaction. The incorporation of three peripheral fins improves final fixation avoiding micromotion at the immediate post operative period. From 1992 we have implanted 4068 cups, the majority of them in primary cases (78%). The rest 22% in revision cases.

Cup characteristics: hemispheric with fins to improve prumary fixation and HA coating; ring-long ystem in common with other Biomet models, reason why liner are interchangeables.

Material and methods.- 4.068 Bihapro cups (Biomet-Merck) were implanted at a multicenter study in three Hospitals, adjoined to the University, between 1992 and 2003. This is a press-fit model with a porous surface coated with HA and three peripheral fins to improve primary fixation and also dome holes to allow the use ob bone screws.

Prymary indication: osteoarthritis (76%), AVN (7%), fractures (8%), dysplasias (3%), rheumatoid arthritis (6%). Surgical approach: lateral (49%), posterolateral (34,2%), anterior (16,8%). Prophylaxis: antitrombotic (LMWH), antibiotic (1st generation cephalosporins), heterotopic ossification (indomethacin).

Results.- Results. 24 patients showed dislocation and 47 % had some degree of periarticular ossification one year alter surgery; the approach used did not show significative differences. The survival study was done using Kaplan-Meier’s curve. The end-point for failure in this study was the need to perform aesptic revision surgery; being the survival at 9 years of 99.49 % (CI 95 % 99.08 – 99.90). Seven cases needed revision surgery (0.3 %); two cases for migration of the cup and five cases for iterative dislocations.

Conclusions.- Acetabular cups with Plasma Spray Porous Coating in combination with HA, results stable at mid term. The supplementary fixation of the three peripheral fins avoids micromotion optimizing long-term fixation.

Theses abstracts were prepared by Professor Roger Lemaire. Correspondence should be addressed to EFORT Central Office, Freihofstrasse 22, CH-8700 Küsnacht, Switzerland.