header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

CYCLIC LOADING OF ROTATOR CUFF REPAIRS: AN IN VITRO BIOMECHANICAL COMPARISON OF BIOABSORBABLE ANCHORS WITH TRANSOSSEOUS SUTURES



Abstract

Introduction and Aims: Suture anchors allow consistent reattachment of tendons and ligaments to bone. Many options are available. The purpose of this study was to compare the initial strength of two rotator cuff repair techniques. The hypothesis was that rotator cuff repair strength with anchors would be inferior to transosseous sutures.

Method: Eight paired shoulders with a standardised supraspinatus defect were randomised to bioabsorbable nonsuture-based anchor or transosseous suture repair. Each specimen was then subjected to a stepwise cyclic loading protocol, utilising a custom-designed loading apparatus. Repair site migration was measured using an optical measurement system, consisting of a digital camera and custom software. Mode of failure, number of cycles and load to failure were measured for 50% (5 mm) and 100% (10 mm) loss of repair. These results were correlated with bone mineral density, age and gender. Statistical analysis utilised paired t-tests and Pearson correlations.

Results: The anchors failed at the anchor-tendon interface, whereas the sutures failed through the sutures. Mean values for 50 percent loss of repair were 206 ± 88 cycles and 44 ± 15 N for the sutures, and 1193 ± 252 cycles and 156 ± 20 N for the anchors (p< 0.05). The corresponding values for 100 percent loss of repair were 2458 ± 379 cycles and 294 ± 27 N for the sutures, and 2292 ± 333 cycles and 263 ± 28 N for the anchors (p> 0.05). These results may be due to the relative less deformability of the anchors. This may be relevant clinically, as in the early post-operative period, while tendon healing to bone is occurring, anchors may offer improved strength, allowing improved initial healing. Strength was unaffected by bone quality. This may be attributed to each repair failing primarily through the repair construct or at the anchor-tendon interface, and not through bone. Strengths of this study include the use of paired specimens, the stepwise cyclic loading protocol, as well as increased accuracy of our measurement system. Limitations include the use of an in vitro model, as well as a simulated, standardised rotator cuff tear.

Conclusion: Repair strength with anchors was superior to sutures. Strength was unaffected by bone quality. Anchors facilitate an arthroscopic procedure, decrease operative time, and may allow a faster post-operative recovery. This study has described a new high-resolution method of measuring tendon repair failure and may be useful in future studies.

These abstracts were prepared by Editorial Secretary, George Sikorski. Correspondence should be addressed to Australian Orthopaedic Association, Ground Floor, The William Bland Centre, 229 Macquarie Street, Sydney, NSW 2000, Australia.

At least one of the authors is receiving or has received material benefits or support from a commercial source.