OPTIMISATION OF RADIOGRAPHIC ANALYSIS OF ACETABULAR POSITION OF HIP PROSTHESES: PRACTICAL DEDUCTIONS FROM 46 CASES



Abstract

Purpose: Uncertain position of the acetabular implant has been the cause of dysfunction in certain cases of total hip arthroplasty (THA). Classical computed tomographic analysis of anteversion has certain limitations. Integrated reconstruction of positions at risk allows a better diagnostic approach.

Material and methods: We studied 46 THA because of posterior malposition (n=17, anterior subluxation in the standing position in twelve, and true dislocation in five) and anterior malposition (n=29, posterior subluxation in sixteen and true dislocation in thirteen). Two groups of 70 naïve hips and a group of 56 THA with no functional problem served as controls. The position of the acetabulum was studied on optimised computed tomography slices reconstructing the planes of analysis for the standing, sitting and reclining positions. The reference planes for the slices was given by the sacral tilt angle measured on the lateral views of the patient in the corresponding positions. The optimised computed tomographic measurements of anteversion were compared with the classical measures. None of the patients had abnormal femoral anteversion and/or an oblique pelvis and/or leg length discrepancy greater than 10 mm. The frontal inclination of the acetabular implants was 40°–50°.

Results: In the naïve hips, acetabular anteversion varied: 19.2 with the conventional method, 15.7 in the standing position and 31 in the sitting position. In the THA controls, anteversion measurements differed: 21.3 with the conventional method, 21.4 in the standing position and 35.8 in the sitting position. In the THA with a posterior malposition, 18/29 could not be explained by the conventional measurement, but the optimised measurement enabled an understanding in 17 hips (defective anteversion in the sitting position).

Discussion: Changes in pelvis orientation between the sitting and standing positions modifies real anteversion of the cup. In particular, subjects with THA tend to have a spontaneous posterior tilt of the pelvis related to trunk ageing. This element should be taken into account for the analysis of both major and minor THA dysfunction.