ACUTE BIOMECHANICAL EFFECTS OF A NOVEL INTRA-DISCAL DECOMPRESSION CATHETER ON HUMAN LUMBAR DISCS



Abstract

Introduction: Intra-Discal Electrothermal Therapy (IDET) has been used to treat chronic discogenic low back pain. A novel intradiscal decompression catheter has been developed to reduce local disc bulging in cases of contained prolapse. This new catheter is inserted percutaneously into a disc and advanced under radiographic control into a postero-lateral position targeting the herniation. The decompression catheter uses more focused heating and higher temperatures than previous devices and is intended to provide a local decompression of the disc through a thermally mediated reduction in nuclear volume. The purpose of this study was to investigate changes in internal stress profiles following use of the new catheter.

Methods: Five cadaveric lumbar ‘motion segments’ were dissected from two spines (age 64–84 yrs). Each segment was compressed, normally to 1 kN, while a miniature pressure transducer was withdrawn from posterior to anterior across the mid-sagittal diameter of the disc producing a baseline stress profile. A decompression catheter was inserted into the disc and its position confirmed with plain radiography. The temperature of the catheter was increased to 90°C over a period of 14 minutes. Stress profiles were then repeated.

Results: Stress profiles in three of the five segments showed changes consistent with degenerative change. In these discs stress profiles following ‘treatment’ showed up to a 35% reduction in the magnitude of stress peaks in the posterior annulus. There was very little change in the distribution of stress in the two non-degenerate discs. Stress in the nucleus appeared unchanged in all discs.

Conclusions: Treatment of degenerate discs with the decompression catheter lead to a measurable alteration in annular stress peaks associated with degenerative discs, while non-degenerate discs were unaffected. These preliminary findings of an ongoing study suggest that the novel decompression catheter has a biomechanical effect in certain classes of disc.