FAILURE OF LONG PROXIMAL FEMORAL NAIL IN A PATHOLOGICAL FRACTURE. A CASE REPORT.



Abstract

A 52 year old male presented with a pathological subtrochanteric femoral fracture secondary to multiple myeloma. While stabilising the fracture with a Long Proximal Femoral Nail (PFN) distal femur fracture occurred, while introducing the distal locking screw, which was fixed with two cables. Partial weight bearing was allowed for the first six weeks.

Three months after surgery the distal static locking screw broke. Eighteen months post surgery patient developed sudden spontaneous right hip pain and was treated with further chemotherapy and radiotherapy. Radiographs showed the fracture had not healed but there was no evidence of implant failure. Two years later patient presented with sudden increase in right hip pain with inability to walk. Radiographs showed that the nail had broken at the proximal hip screw hole.

At revision surgery, with difficulty the broken distal locking screws were removed and the broken nail was removed by pushing it from below through the knee. The non union was stabilised with another long PFN. At four months post revision surgery there were radiological signs of bone healing and patient had no symptoms.

Discussion: Reconstruction nails such as long PFN are bio mechanically suited for proximal femoral fractures and metastases. Bone cement augmentation has been reported to provide additional support in metastases. Dynamisation of the fracture leads to fracture impaction and promotes fracture healing. In this case implant failure was probably due to non union and fatigue failure of the implant. In spite of ‘spontaneous’ dynamisation (broken static distal screw), union did not occur initially.

This is the first reported incidence of failure of long PFN in a pathological femoral fracture stabilisation.

The abstracts were prepared by Mr Roger Tillman. Correspondence should be addressed to BOOS at the Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PN