We analysed the long-term clinical and radiological results of 63 uncemented Low Contact Stress total knee replacements in 47 patients with rheumatoid arthritis. At a mean follow-up of 12.9 years (10 to 16), 36 patients (49 knees) were still alive; three patients (five knees) were lost to follow-up. Revision was necessary in three knees (4.8%) and the rate of infection was 3.2%. The mean clinical and functional Knee Society scores were 90 (30 to 98) and 59 (25 to 90), respectively, at final follow-up and the mean active range of movement was 104° (55° to 120°). The survival rate was 94% at 16 years but 85.5% of patients lost to follow-up were considered as failures. Radiological evidence of impending failure was noted in one knee.


Correspondence should be sent to Mr S. Sharma at 47 Shuna Place, Newton Mearns, Glasgow G77 6TN, UK; e-mail: sssharma@hotmail.com