header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

O3078 MENISCOFEMORAL LOADING AFTER REPAIR OF BUCKET HANDLE TEARS WITH BIODEGRADABLE IMPLANTS: A BIOMECHANICAL STUDY



Abstract

Aims: To clarify whether joint loading after meniscus repair causes increased pressure on the femoral condyle and is responsible for early cartilage damage. Methods: In sixteen human cadaver knees a bucket handle tear was created at the posterior horn of the medial and lateral meniscus. The lesion was repaired using two biodegradable implants (either Stingerª, Arrow¨, Dartª or Meniscal Screwª) on each occasion. Loading was compared to intact menisci and menisci after suture repair using 2/0 Ethibond. The specimens were mounted into specially designed jigs, taking all degree of freedom of the knee joint into account, and þxed to a material testing machine (Bionix 858 MTS). Constant loading of 350 N was performed during knee motion of 0¡Ð90¡ of ßexion. The meniscofemoral pressure was measured using the Tekscansystem. All þxation techniques were tested þve times. Results: Increased joint loading at the posterior horn occurred with increased ßexion angle of 0¡, 30¡, 60¡ and 90¡ of knee ßexion in the medial and lateral compartment (p< 0.05). No signiþcant increase in joint loading was noticed after meniscus repair with biodegradable implants. Conclusions: Biodegradable implants do not cause higher meniscofemoral joint loading due to meniscus implants in the posterior horn and resulting cartilage damage at the femoral condyle is unlikely.

Theses abstracts were prepared by Professor Dr. Frantz Langlais. Correspondence should be addressed to him at EFORT Central Office, Freihofstrasse 22, CH-8700 Küsnacht, Switzerland.