SURGICAL REVISION AFTER FAILURE OF THE GLENOID COMPONENT OF SHOULDER PROSTHESES: 16 CASES



Abstract

Purpose: Failure of the glenoid component is the main complication of total shoulder prostheses. When surgical revision is necessary, the surgeon has the option of a new implantation or non-prosthetic plasty (glenoido-plasty). The purpose of the present work was to analyse results obtained with these two techniques in order to propose proper indications.

Material and methods: This retrospective study included 16 patients, mean age 62 years at revision surgery. Fialures included loosening of a cemented glenoid implant (n=9) and failure of non-cemented implants (3 defective anchors, 4 unclipped polyethylene inserts). Mean time to revision was 39 months (2–178) after primary implantation. A new glenoid implant was cemented in nine patients (group A). Seven patients (group 2) had glenoidoplasty with an iliac graft in four.

Results: A mean follow-up of 37 months (19–73), the Constant score had improved from 18 points before revision to 52 points (+34). Two patients experienced a complication requiring a second revision (infection, instability) and one patient underwent subsequent surgery for biceps tenodesis. For the group with glenoidoplasty with iliac graft, insertion of an inverted prosthesis was achieved during a second operative time. In group 1, the mean Constant score at last-follow-up was 63 points (+45) with the pain score of 11, movement score of 29. In group 2, the mean Constant score was 37 (+19) with pain at 6 and motion at 16. In this group, the mean score was 48 points with a glenoid graft and 21 points with simple implant replacement.

Discussion: Revision surgery for a failed glenoid implant remains a difficult procedure but can be effective for pain relief and improved motion score. The small number of patients in this series makes it difficult to perform statistical analysis but the results do point in favour of prosthetic reimplantation when the bone stock is sufficient. For other patients, a graft would be preferable to simple implant removal. This would allow secondary revision if possible.

The abstracts were prepared by Docteur Jean Barthas. Correspondence should be addressed to him at Secrétariat de la Société S.O.F.C.O.T., 56 rue Boissonade, 75014 Paris.