header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

A SEMIACTIVE ROBOT FOR PRECISE RESECTION BY MILLING IN TKA



Abstract

The disadvantages of sawing for precise bone cuts are well known: untrue cuts, heat and metal wear. The main limiting factors of available milling devices are the difficult handling and high costs, especially if the devices are based on a robot. Supported by clinical users and mechanical engineers a milling concept adopted from machining has been realised in order to overcome this limitations. The „All-in-One Milling-Tool“ achieves the same precision of a robot by a mechanically guided milling resection far below the necessary investment for a robot. Three methods are provided for the alignment of the resection planes and will be discussed: intramedullary adjustment, 3D CT-based planning and intramedullar performance as well as the performance under control by navigation. All versions are based on a handheld resection and guarantee a visual and haptical feedback for the surgeon. The use of navigation has the advantage of the accurate transfer of the 3D plan into the OR, the interactive facilitated alignment und resection steps and the documentation of planned and actual implant position.

The abstracts were prepared by Nico Verdoschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, Universitair Medisch Centrum, Orthopaedie / CSS1, Huispost 800, Postbus 9101, 6500 HB Nijmegen, Th. Craanenlaan 7, 6525 GH Nijmegen, The Netherlands.